General Measurement Considerations
C-3
Figure C-2
Eliminating ground loops
Triboelectric effects
Triboelectric currents are generated by charges created between a conductor and an insula-
tor due to friction. Here, free electrons rub off the conductor and create a charge imbalance
that causes the current flow. For example, bending a triaxial cable causes friction between
the center conductor (HI) and its surrounding insulator resulting in triboelectric currents.
Triboelectric currents can be minimized as follows:
•
Use “low noise” cables. These cables are specially designed to minimize charge
generation and use graphite to reduce friction. The Keithley Model 4801–4803
cables are low noise.
•
Use the shortest cables possible, and secure them (i.e., taping or tying) to a non-vibrating
surface to keep them from moving.
Piezoelectric and stored charge effects
Piezoelectric currents are generated when mechanical stress is applied to certain insulating
materials (i.e., crystalline). In some plastics, pockets of stored charge cause the material to
behave in a similar manner.
When building test fixtures, choose good insulating materials and make connecting struc-
tures as rigid as possible. Make sure there are no mechanical stresses on the insulators.
Instrument 1
Instrument 2
Instrument 3
Power Line Ground
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......