Model 6485 Picoammeter Instruction Manual
Measurement Concepts
2-13
Figure 2-8
Floating measurements
Zero check and zero correct
lists the display messages associated with zero check and zero correct. The
two-character message is displayed along with the reading.
Zero check
When zero check is enabled (on), the input amplifier is reconfigured to shunt the input sig-
nal to low as shown in
From the front panel, enable / disable zero check by pressing the ZCHK key. Refer to
for bus commands.
Leave zero check enabled when connecting or disconnecting input signals.
Table 2-3
Display messages for zero check and zero correct
Display
Message
Zero Check Zero Correct
ZC
On
Off
ZZ
On
On
CZ
Off
On
20V
R
1
20V
A
+
–
R
2
6485
Picoammeter
R
3
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......