3-12
Measurements and Sourcing Voltage
Sourcing voltage
CAUTION
Do not connect external sources to the 6487 voltage source. External
sources may damage the 6487 voltage source.
Turning source on and off
The OPER key will toggle the voltage source state (on/off). The Model 6487 will output
the configured voltage, and the VOLTAGE SOURCE OPERATE indicator will turn on.
OPER (operate) key
The OPER (operate) key will function to turn the voltage source off, even if the instrument
is operating under remote control (REM annunciator on), assuming that the LLO (Local
Lockout) function has not been employed. While in remote, the OPER key will only turn
the source off. To turn it on, the Model 6487 must be in local (see
).
Voltage source off state
The voltage source is not in a high-impedance state when it is turned off. Rather, it is in a
state that acts just like the voltage source was programmed to 0V on the selected range. It
will enter this state on power-up after the VOLTAGE SOURCE OPERATE light blinks. In
contrast, the safety interlock will cause the voltage source to go into a high-impedance
state instead of 0V output, and the source will stay in the high-impedance state until the
operate state is changed to on. The exception is the 10V range where the interlock is
optional. The OPERATE light and front panel display do not indicate the difference
between 0V output and high-impedance output caused by an open interlock. The interlock
status is available by query via remote (see
Compliance indication
At any time, it is possible that the voltage source will go into compliance (current limit
reached). Should this situation occur, the OCOMP annunciator (Output Compliance) will
flash, and the displayed voltage value (visible when current readings are configured for
readings of less than 6
½
digits) will alternate between showing the value and displaying
“CMPL”. If you are in a menu where the voltage source value is not shown on the right-
most four characters of the display, only the flashing OCOMP annunciator will be shown.
Table 3-6
Voltage source current limits
Source Range:
Selectable Current Limit
10.0000V Range
25µA
250µA
2.5mA
25mA
50.000V Range
25µA
250µA
2.5mA
500.00V Range
25µA
250µA
2.5mA
Summary of Contents for 6487
Page 13: ......
Page 15: ......
Page 35: ...1 20 Getting Started Model 6487 User s Manual...
Page 47: ...2 12 Connections Model 6487 User s Manual...
Page 61: ...3 14 Measurements and Sourcing Voltage Model 6487 User s Manual...
Page 85: ...7 8 Remote Operation and Commands Model 6487 User s Manual...
Page 86: ...A Specifications...
Page 88: ...B GeneralMeasurement Considerations...
Page 94: ...C ExamplePrograms...
Page 101: ......
Page 103: ......
Page 104: ......