Measurements and Sourcing Voltage
3-13
Open interlock indication
If the interlock is asserted (opened) while the unit is on the 50V or 500V range, the voltage
source will also technically be in compliance. However, there will be no indication of that
status over the front panel or in the status registers. The open interlock takes precedence.
SCPI commands — voltage source
Programming example
— voltage
The following command sequence will output 5V on the 10V range with a 2.5mA limit:
*RST
' Return 6487 to GPIB defaults.
SOUR:VOLT:RANG 10
' Select 10V source range.
SOUR:VOLT 5
' Set voltage source output to 5.
SOUR:VOLT:ILIM 2.5e-3
' Set current limit to 2.5mA.
SOUR:VOLT:STAT ON
' Put voltage source in operate.
Table 3-7
SCPI commands — voltage source
Command
Description
SOUR:VOLT <Voltage>
SOUR:VOLT:RANG <Range>
SOUR:VOLT:ILIM <Limit>
SOUR:VOLT:STAT <b>
SOUR:VOLT:INT <b>
SOUR:VOLT:INT:FAIL?
Set the source output level (-500 to +500V).
Set source range to one of three ranges: 10, 50, and 500V.
Set the source current limit: 2.5e-5, 2.5e-4, 2.5e-3, or 2.5e-2.
1
Enable (ON) or disable (OFF) source output.
Enable (ON) or disable (OFF) interlock for 10V range.
2
Query interlock state (1 = asserted); source output cannot be turned on.
1. 2.5e-2 not available for 50V and 500V ranges.
2. See
Summary of Contents for 6487
Page 13: ......
Page 15: ......
Page 35: ...1 20 Getting Started Model 6487 User s Manual...
Page 47: ...2 12 Connections Model 6487 User s Manual...
Page 61: ...3 14 Measurements and Sourcing Voltage Model 6487 User s Manual...
Page 85: ...7 8 Remote Operation and Commands Model 6487 User s Manual...
Page 86: ...A Specifications...
Page 88: ...B GeneralMeasurement Considerations...
Page 94: ...C ExamplePrograms...
Page 101: ......
Page 103: ......
Page 104: ......