4-2
Range, Units, Digits, Rate, and Filters
Range, units, and digits
Range
The ranges for current measurements are listed in
.
The full scale readings for every measurement range are 5% over range. For example, on
the 20
µ
A range, the maximum input current is ± 21
µ
A. Input values that exceed the maxi-
mum readings cause the overflow message OVRFLOW to be displayed.
Manual ranging
To select a range, press the manual RANGE key. The instrument changes one range per
key-press. If the instrument displays the OVRFLOW message on a particular range, select
a higher range until an on-range reading is displayed. Use the lowest range possible with-
out causing an overflow to ensure best accuracy and resolution.
Autoranging
When using autorange, the instrument automatically goes to the most sensitive available
range to measure the applied signal. Up-ranging occurs at 105% of range, while
down-ranging occurs at the range value. The AUTO key toggles the instrument between
manual ranging and autoranging. The AUTO annunciator turns on when autoranging is
selected. To disable autoranging, press AUTO or a manual RANGE key. Pressing AUTO
to disable autoranging leaves the instrument on the present range.
Every time an autorange occurs, a search for every available range of the selected function
is performed. The time it takes to perform the search could slow down range change speed
significantly. Setting upper and/or lower autorange limits can reduce search time.
Autorange limits
Search time for finding the correct amps range can be reduced by setting upper and/or
lower autorange limits. To set upper and/or lower autorange limits, press CONFIG, then
either manual RANGE key (for UPPER or LOWER), then scroll through the available
range limits using the RANGE keys. Press ENTER when the desired range is flashing.
Table 4-1
Measurement ranges
nA
µ
A
mA
2nA
2
µ
A
2mA
20nA
20
µ
A
20mA
200nA
200
µ
A
Summary of Contents for 6487
Page 13: ......
Page 15: ......
Page 35: ...1 20 Getting Started Model 6487 User s Manual...
Page 47: ...2 12 Connections Model 6487 User s Manual...
Page 61: ...3 14 Measurements and Sourcing Voltage Model 6487 User s Manual...
Page 85: ...7 8 Remote Operation and Commands Model 6487 User s Manual...
Page 86: ...A Specifications...
Page 88: ...B GeneralMeasurement Considerations...
Page 94: ...C ExamplePrograms...
Page 101: ......
Page 103: ......
Page 104: ......