10-18
Status Structure
Model 6487 Reference Manual
Output queue
The output queue holds data that pertains to the normal operation of the instrument. For
example, when a query command is sent, the response message is placed in the output
queue.
When data is placed in the output queue, the message available (MAV) bit in the status
byte register sets. A data message is cleared from the output queue when it is read. The
output queue is considered cleared when it is empty. An empty output queue clears the
MAV bit in the status byte register.
A message is read from the output queue by addressing the Model 6487 to talk after the
appropriate query is sent.