14-14
SCPI Reference Tables
Model 6487 Reference Manual
:DISable <list>
Specify messages not to be placed in queue.
(Note 5)
:DISable?
Read the disabled messages.
:CLEar
Clear messages from error queue.
Parameters:
<NDN>
= #Bxx…x
Binary format (each x = 1 or 0)
= #Hx
Hexadecimal format (x = 0 to FFFF)
= #Qx
Octal format (x = 0 to 177777)
<NRf>
= 0 to 65535
Decimal format
<list>
= (100:200, -224)
Example of a range and single entry (+100 t200 and -224).
Notes:
1. Commands in this subsystem are not affected by *RST or SYSTem:PRESet. The effects of cycling power, *CLS, and
STATus:PRESet are explained by the following notes.
2. Event registers — Power-up and *CLS clears all bits. STATus:PRESet has no effect.
3. Enable registers — Power-up and STATus:PRESet clears all bits. *CLS has no effect.
4. Error queue — Power-up and *CLS empties the error queue. STATus:PRESet has no effect.
5. Error queue messages — Power-up enables error messages and disables status messages. *CLS and STATus:PRESet have no
effect.
Table 14-6 (cont.)
STATus command summary
Command
Description
Default
parameter
Ref
Section SCPI