G-14
Applications Guide
Model 6487 Reference Manual
Up-range input response
illustrates the type of transient voltage that can be expected when up-ranging
with a full-scale input signal (200
μ
A signal on 200
μ
A range, up-range to 2mA range).
Both the magnitude and duration of this voltage are reduced for lower current ranges. The
current limit imposed by the R
F
is also greatly reduced. The polarity depends on the polar-
ity of the input current.
was measured with a positive input current.
Figure G-11
Transient Voltage
When it is necessary to up-range during auto-ranging operation, multiple ranges may be
crossed to find the correct range. The duration of the transient in
extended in this case, but the magnitude will not increase significantly.