Model 6487 Reference Manual
Performance Verification
15-11
5.
Verify voltage source accuracy for each of the values listed in
. For each
test point:
•
Select the correct Model 6487 voltage source range.
•
Set the voltage source output to the indicated value.
•
Make sure the voltage source is in operate (output on).
•
Verify that the DMM voltage reading is within the limits given in the table.
6.
Repeat the procedure for negative source currents with the same magnitudes as
those listed in
.
Figure 15-3
Connections for voltage source output accuracy
Table 15-4
Reading limits for voltage source accuracy
Model 6487
Source Range
Output
Voltage
Model 6487 Output Voltage Limits
(1 Year, 18°C-28°C)
10V
10.000V
9.989 to 10.011V
50V
50.000V
49.946 to 50.054V
500V
500.00V
499.21 to 500.79V
NEXT
PREV
POWER
DISPLAY
2001
MULTIMETER
DCV
ACV
DCI
ACI
Ω
2
Ω
4
FREQ
TEMP
REL
TRIG
INFO
LOCAL
EXIT
ENTER
RANGE
RANGE
!
F
R
500V
PEAK
FRONT/REAR
2A 250V
AMPS
CAL
STORE RECALL
CHAN
SCAN
FILTER MATH
CONFIG MENU
HI
INPUT
LO
SENSE
Ω
4 WIRE
INPUTS
3
50V
PEAK
1100V
PEAK
AUTO
Model 2001 DMM
120
LINE RATING
50, 60Hz
50 VA MAX
INPUT
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
IEEE-488
CAT
I
TRIGGER LINK
RS-232
MADE IN
U.S.A.
ANALOG OUT
DIGITAL I/O
505V
MAX
V-SOURCE OUTPUT
505V PK TO CHASSIS
LO
HI INTERLOCK
!
!
FUSE
LINE
400mAT
(SB)
100 VAC
120 VAC
200mAT
(SB)
220 VAC
240 VAC
505V PK
505V PK
505V PK
Model 64
8
7
I
N
PUT
V-
S
OURCE
OUTPUT