4-12
Range, Units, Digits, Rate, and Filters
Model 6487 Reference Manual
SCPI programming — filters
Programming example
The following command sequence configures and enables both filters:
' Median Filter:
MED:RANK 5
' Set rank to 5.
MED ON
' Enable median filter.
' Digital Filter:
AVER:COUN 20
' Set filter count to 20.
AVER:TCON MOV
' Select moving filter.
AVER ON
' Enable digital filter.
Table 4-6
SCPI commands — filters
Commands
Description
Default
For median filter:
[:SENSe[1]]
:MEDian
:RANK <n>
[:STATe] <b>
For digital filter:
[:SENSe[1]]
:AVERage
:TCONtrol <name>
:COUNt <n>
[:STATe] <b>
SENSe Subsystem:
Median Filter:
Specify filter rank: 1 to 5.
Enable or disable median filter.
SENSe Subsystem:
Digital Filter:
Select filter control: MOVing or REPeat.
Specify filter count: 2 to 100.
Enable or disable digital filter.
1
OFF
MOV
10
OFF