Model 6514 Quick Results Guide
Basic operation
Selecting operating modes
Most front panel keys have a dual operation or function. The nomenclature on a key
indicates its primary or unshifted function or operation. The nomenclature above a key
indicates its shifted function or operation. To select a primary function press the
corresponding key. To select a shifted function, press SHIFT first and then the desired key.
For example, press AVG to select the digital filter. Press SHIFT then MX+B to choose the
mX + b function.
Making volts and ohms measurements
The basic procedure to make volts and ohms measurements is provided in Table 1.
Table 1
Volts and ohms measurement procedure
Procedure
Details
1. Enable zero check.
Press ZCHK to enable zero check before making
connection changes or selecting function.
2. Select function.
Press V to measure voltage.
Press
Ω
to measure ohms.
3. Enable or disable guard.
Use the GRD key to select depending on guarded
or unguarded connections.
4. Perform zero correct (volts only).
Press ZCOR to zero correct 6514 for maximum
volts accuracy.
5. Select range.
Use RANGE
5
and
6
keys to manually
select range. Or press AUTO RANGE (AUTO
annunciator on) to enable auto range.
6. Connect DUT to 6514 input.
See Figure 4 (unguarded) or Figure 5 (guarded).
7. Disable zero check.
Press ZCHK to disable zero check in order to
make measurements.
8. Take readings from display.
Observe the volts or ohms reading on the front
panel display.