Section 7: Triggering
Model 6517B Electrometer Reference Manual
7-2
6517B-901-01 Rev. C / August 2015
Trigger configuration menu
The Keithley Instruments Model 6517B Electrometer triggers are set up from the
CONFIGURE
TRIGGER
menu. The menu structure is shown and summarized in the following table.
Notice that there are two trigger configuration structures: BASIC and ADVANCED. The basic menu
structure can be used when simple trigger operations suffice. The advanced menu structure must be
used when more sophisticated trigger operations (such as scanning) are required. The differences
between basic and advanced triggering are explained below.
Trigger configuration menu
Menu item
Description
BASIC
MODE
CONTINUOUS
ONE-SHOT
SOURCE
IMMEDIATE
MANUAL
GPIB
EXT
TIMER
Select and configure basic triggering:
Select trigger mode:
Use for continuous triggering
Use for one-shot triggering
Select source of triggers:
Use to make measurements immediately
Use TRIG key to control measuring
Use bus triggers to control measuring
Use external triggers to control measuring
Use a timer to control measuring; enter trigger interval
(0.001 - 999999.999 seconds)
ADVANCED
MEASURE
SOURCE
IMMEDIATE
EXTERNAL
MANUAL
GPIB
TRIGLINK
TIMER
HOLD
DELAY
COUNT
INFINITE
ENTER-CHAN-COUNT
CONTROL
SOURCE
ACCEPTOR
Select and configure advanced triggering:
Measure layer menu:
Select measure source:
Use to make measurements immediately
Use external trigger to control measuring
Use TRIG key to control measuring
Use bus triggers to control measuring
Use trigger link triggers to control measuring; enter trigger link
mode and lines
Use a timer to control measuring and enter interval between
triggers (0.001 - 999999.999 seconds)
Use to hold up the measurement in the measure layer
Use to delay measurement in the measure layer
(0.001 - 999999.999 seconds)
Define number of measurements to make:
Repeat measuring indefinitely
Specify count (1 - 99999)
Select trigger control mode:
Enable source Bypass
Disable source Bypass