Model 6517B Electrometer Reference Manual
Section 14: SCPI command reference
6517B-901-01 Rev. C / August 2015
14-87
M8009:RSWitch?
This command is used to query the switch setting on the Model 8009.
Usage
[:SENSe[1]]:RESistance:RESistivity:M8009:RSWitch?
Details
When using the Model 8009 Resistivity Test Fixture, the Model 6517B senses the switch setting
(Surface or Volume) of the test fixture through the safety interlock cable and automatically configures
the instrument for that resistivity measurement type.
This query command is used to read the switch setting (Surface or Volume) of the Model 8009
Resistivity Test Fixture.
Also see
None
:USER commands
The following commands are used to configure the instrument for resistivity measurements when
using a user-supplied test fixture. These command are not needed if using the Model 8009 Resistivity
Test Fixture.
:RSELect <name>
This command is used to specify the measurement type for the user test fixture.
Usage
[:SENSe[1]]:RESistance:RESistivity:USER:RSELect <name>
:RSELect?
<name>
SURFace:
Surface resistivity
VOLume:
Volume resistivity
Details
This command is used to specify the resistivity type (
SURFace
or
VOLume
) for the user-supplied test
fixture. This command is not needed when using the Model 8009 test fixture.
Also see
None