Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-132
6517B-901-01 Rev. C / August 2015
:HSControl <b>
This command controls humidity readings.
Usage
:SYSTem:HSControl <b>
:HSControl?
<b>
0
or
OFF
: Disable humidity readings
1
or
ON
: Enable humidity readings
Details
This command controls humidity readings. When enabled (and the Model 6517-RH connected to the
instrument), the Model 6517B makes humidity readings. Note that
HUMidity
has to be a selected
data element in order to include the humidity reading in the data string.
Also see
None
:HLControl <b>
This command controls the A/D hardware limit.
Usage
:SYSTem:HLControl <b>
:HLControl?
<b>
0
or
OFF
: Disable A/D hardware limit
1
or
ON
: Enable A/D hardware limit
Details
This command is used to control the A/D hardware limit. With the hardware limit
ON
, the "OutOfLimit"
error message is enabled to indicate that an out of range noise spike has occurred.
Disabling the hardware limit disables the error message. With the hardware limit
OFF
, noise spikes
may cause the reading to be erroneous.
Also see
None
RS-232 interface commands
The following commands are provided for control (enabled or disabled) of front-panel keys while
using the RS-232 interface.