Section 14: SCPI command reference
Model 6517B Electrometer Reference Manual
14-136
6517B-901-01 Rev. C / August 2015
:TIMer <n>
This command sets the timer interval.
Usage
:SYSTem:MACRo:TRIGger:TIMer <n>
<n>
0.001 to 99999.999:
Specify timer interval in seconds
DEFault:
0.1 seconds
MINimum:
0.001 seconds
MAXimum:
99999.999 seconds
Details
This command is used to specify the time interval for the timer. The
:SOURce TIMer
command is
used to select the timer as the control source.
Also see
None
:INTerlock?
This command determines if the safety interlock cable is connected to the Model 6517B.
Usage
SYSTem:INTerlock?
Details
This query command is used to determine if the safety interlock cable is connected to the Model
6517B. The response message is interpreted as follows:
1
= Interlock cable connected to the 6517B and fixture, or not connected to 6517B at all.
0
= Interlock cable connected to 6517B but not connected to fixture, or interlock cable connected to
6517B but fixture lid is open.
If the interlock cable is not connected, the Model 6517B cannot determine the state of the test fixture
lid (open or closed) or read the test fixture switch settings. A potential safety hazard is present when
the fixture lid is open.
Also see
None
:TRACe subsystem
The commands in this subsystem are used to configure and control data storage into the buffer. The
commands are summarized in :
TRACe command summary
(on page 14-25). See Buffer (data store)
for more information about the buffer.