Section 4: Basic measurements
Model 6517B Electrometer Reference Manual
4-12
6517B-901-01 Rev. C / August 2015
Amps configuration
The following information explains the various configuration options for the amps function. The
configuration menu is summarized in the following table. This menu is accessed by pressing
CONFIG
and then
I
.
Note that a function does not have to be selected in order to be configured. When the function is
selected, it assumes the pre-programmed status.
Amps configuration menu
Menu item
Description
SPEED
NORMAL
FAST
MEDIUM
HIACCURACY
SET-SPEED-EXACTLY
SET-BY-RSLN
Measurement speed (integration time) menu:
Select 1 PLC (power line cycle, 16.67 ms for 60Hz, 20 ms for 50 Hz and 400 Hz)
Select 0.01 PLC
Select 0.1 PLC
Select 10 PLC
Set integration in PLC (0.01-10)
Default to setting appropriate for resolution
FILTER
AVERAGING
TYPE
NONE
AVERAGING
ADVANCED
AVERAGING-MODE
MEDIAN
DISABLE
ENABLE
Filter menu:
Configure digital averaging filter:
Select type of average filter:
No average filtering performed
Program a simple average filter (1-100 readings)
Program a simple average filter (1-100 readings) with noise tolerance window
(0-100 % of range)
Select moving average or repeating average mode
Configure median filter:
Disable median filter
Enable median filter and specify rank (1-5)
RESOLUTION
AUTO
3.5 d, 4.5 d, 5.5 d, 6.5 d
Display resolution menu:
Default to resolution appropriate for integration time
Select a specific resolution
AUTO-RANGE
USE-ALL-RANGES
SET-LIMITS
MIN-AUTO
MAX-AUTO
Autorange menu:
Use all ranges when autoranging
Limit the ranges used in the autorange search:
Specify the minimum range in the search
Specify the maximum range in the search
DAMPING
Enable or disable damping
SPEED
The SPEED parameter sets the integration time of the A/D converter, the period of time the input
signal is measured (also known as aperture).
FILTER
Use this menu item to configure the two basic filter types: averaging and median. Note that you can
use either the averaging filter, the median filter, or both.
The filter menu is available from the function configuration menus (for example, press CONFIG V) or
by pressing CONFIG FILTER with the desired function already selected.