Service Information
5-9
Table 5-3
Channel-to-channel isolation test summary
Test
number
Channel-to-channel isolation
Test equipment location
Channel
closed*
1
2
3
4
5
6
7
8
9
Bank A, Input 1 to Bank A, Input 2
Bank A, Input 2 to Bank A, Input 3
Bank A, Input 3 to Bank A, Input 4
Bank A, Input 4 to Bank A, Input 5
Bank A, Input 5 to Bank A, Input 6
Bank A, Input 6 to Bank A, Input 7
Bank A, Input 7 to Bank A, Input 8
Bank A, Input 8 to Bank A, Input 9
Bank A, Input 9 to Bank A, Input 10
Bank A and Input 1
Bank A and Input 2
Bank A and Input 3
Bank A and Input 4
Bank A and Input 5
Bank A and Input 6
Bank A and Input 7
Bank A and Input 8
Bank A and Input 9
1!2
1!3
1!4
1!5
1!6
1!7
1!8
1!9
1!10
*Assumes Model 7015 installed in slot 1 of mainframe. Programmed as slot (1) and channel.
Figure 5-5
Channel-to-channel isolation test connections
A
B
C
D
1
3
4
5
6
7
8
9
10
2
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
H
L
Bank
Outputs
Bank Inputs
Model 7015
HI
Model 617
INPUT
Model 7025
Unterminated
Triax Cable
(Red)
Banana to Banana Cable
Ground Link
Removed
Unterminated
Banana Cables
Source V and
Measure V/I
Note : Setup shown is configured
to test isolation between
path 1!1 and 1!2.
Jumper
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for 7015-S
Page 88: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 89: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 91: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 92: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...