Card Connections and Installation
3-6
Typical connection scheme
This paragraph shows how the Model 7036 could be
connected to a system that activates external devices and
circuits.
Figure 3-4 shows how external connections can be made to
the system using a custom-built cable that is unterminated at
one end. The unterminated end of the cable can be hard-
wired directly to the external devices and power supplies.
The other end will mate to the Model 7036 switch card. Con-
nection details are provided in the Multi-pin (mass termina-
tion) connections paragraph.
If adding more Model 7036 cards to a system, simply wire
them in the same manner as the first. Remember that
Model 7036 cards installed in the same mainframe are elec-
trically isolated from each other.
Figure 3-4
Typical connection scheme for Model 7036
IN 1
Indicator
Lamp
5V
7036
12V
30V
Alarm
Relay
Equivalent Circuit
Relay
Alarm
5V
12V
30V
Indicator
Lamp
IN 21
IN 2
IN 22
IN 3
IN 23
Custom-built cable
terminated with
7011-KIT-R
7036
OUT 1
OUT 21
OUT 2
OUT 22
OUT 3
OUT 23