Summary of Contents for 7038

Page 1: ...c i a n sw i t ho v e r 1 5 0y e a r so f s p e c i a l i z a t i o n 9 0d a yw a r r a n t y 5d a yr i g h t o f r e t u r no na l l e q u i p me n t 1 3y e a r w a r r a n t i e sf o r n e wa n d p...

Page 2: ...Instruction Manual Model 7038 75 2GHz Multiplexer Card Contains Operating and Servicing Information 7038 901 01 Rev A 6 98...

Page 3: ...LAR USE THE REMEDIES PROVIDED HEREIN ARE BUYER S SOLE AND EXCLUSIVE REMEDIES NEITHER KEITHLEY INSTRUMENTS INC NOR ANY OF ITS EMPLOYEES SHALL BE LIABLE FOR ANY DIRECT INDI RECT SPECIAL INCIDENTAL OR CO...

Page 4: ...Model 7038 75 2GHz Multiplexer Card Instruction Manual 1998 Keithley Instruments Inc All rights reserved Cleveland Ohio U S A First Printing June 1998 Document Number 7038 901 01 Rev A...

Page 5: ...orporate immediately into the manual Addenda are num bered sequentially When a new Revision is created all Addenda associated with the previous Revision of the manual are incorporated into the new Rev...

Page 6: ...voltage is present in any unknown circuit before measuring Users of this product must be protected from electric shock at all times The responsible body must ensure that users are prevented access and...

Page 7: ...result in personal injury or death Always read the associated infor mation very carefully before performing the indicated procedure The CAUTION heading in a manual explains hazards that could damage...

Page 8: ...CONTACT RESISTANCE 1 CONTACT LIFE 3 105 closures 24VDC 10mA DC 1 105 closures 10W 1 2GHz signal 5 106 closures cold switching AC PERFORMANCE 10 100 500 900 2 For ZL ZS 75 MHz MHz MHz MHz GHz Insertion...

Page 9: ...mplified schematic 2 1 2 3 2 Card configuration 2 1 2 3 3 Input output connecting cables 2 3 2 3 4 Typical connecting scheme 2 3 2 4 Card installation and removal 2 4 2 4 1 Card installation 2 4 2 4 2...

Page 10: ...istance tests 4 2 4 3 5 Channel isolation tests 4 2 4 3 6 Special handling of static sensitive devices 4 7 4 4 Principles of operation 4 7 4 4 1 Block diagram 4 7 4 4 2 Relay control 4 7 4 4 3 Switchi...

Page 11: ...Operation Figure 3 1 T connector multiplexer expansion 3 4 Figure 3 2 4 4 matrix multiplexer connections 3 4 Figure 3 3 1 8 multiplexer configuration 3 5 4 Servicing Figure 4 1 Connections for insert...

Page 12: ...nded verification equipment 4 2 Table 4 2 Insertion loss values 4 2 Table 4 3 Recommended troubleshooting equipment 4 8 Table 4 4 Troubleshooting procedure 4 8 5 Replaceable Parts Table 5 1 Model 7038...

Page 13: ...ng the card for repair be sure to fill out and in clude the service form at the back of this manual in order to provide the repair facility with the necessary information 1 4 Manual addenda Any improv...

Page 14: ...s as ordered 2 SMB 10 jumpers CA 193 1A 1 8 Instruction manual If an additional Model 7038 Instruction Manual is required order the manual package Keithley part number 7038 901 00 The manual package i...

Page 15: ...aces components or connector insulators Operate the card in a clean environment If the card be comes contaminated it should be thoroughly cleaned as explained in Section 4 2 3 Connections This paragra...

Page 16: ...configuration Mini SMB Connector Configuration HI LO Signal Common IN1 IN2 IN3 IN4 OUTA Multiplexer A IN5 IN6 IN7 IN8 OUTB Multiplexer B IN9 IN10 IN11 IN12 OUTC Multiplexer C Multiplexer C Output Mul...

Page 17: ...ee Switching considerations in Section 3 for more information 2 3 4 Typical connecting scheme Figure 2 3 shows a typical connecting scheme for the Model 7038 In this arrangement sources are connected...

Page 18: ...me WARNING Turn off power to all instrumentation including the Model 7001 7002 and disconnect all line cords Make sure all power is removed and any stored energy in external circuitry is discharged 1...

Page 19: ...G Turn off power to all instrumentation including the Model 7001 7002 and disconnect all line cords Make sure all power is removed and any stored energy in external circuitry is discharged 1 Pull out...

Page 20: ...bank B channels 5 to 8 Multiplexer bank C channels 9 through 12 NOTE On power up the Model 7001 display will not show that channels 1 5 and 9 are nor mally closed If the instrument is to be op erated...

Page 21: ...2 12 Note that channels are scanned in the order they appear in the scan list To perform a manual scan first select the RESET default conditions in the SAVESETUP selection of the main MENU Press STEP...

Page 22: ...s switched by the card The effects of this characteristic depend on the particular isolation specification Center to shield isolation resistance for example may cause loading problems with signals hav...

Page 23: ...together using short coaxial cables while the remaining output is connected to one of the other multiplexer bank inputs Note that this connection scheme effectively forms a 4 4 matrix with a 4 input m...

Page 24: ...tput Note that IN11 and IN12 are not used with this configuration Aside from providing a 1 8 mux there are a couple other advantages to using this configuration For one thing the characteristic impede...

Page 25: ...not touch areas adjacent to electrical contacts When servicing the card wear clean cotton gloves Do not store or operate the card in an environment where dust could settle on the circuit board Use dr...

Page 26: ...re the signal output cable is connected to the cor rect output jack Close only the channel being tested Verify that the insertion loss values at the various fre quencies are within the limits stated i...

Page 27: ...7 Turn on the Model 6517A voltage source and allow the reading to settle 8 Verify that the Model 6517A resistance reading is 100M 9 Turn off the voltage source and open channel 1 10 Repeat the proced...

Page 28: ...l isolation tests WARNING NO INTERNAL OPERATOR SERVICABLE PARTS SERVICE BY QUALIFIED PERSONNEL ONLY WARNING NO INTERNAL OPERATOR SERVICABLE PARTS SERVICE BY QUALIFIED PERSONNEL ONLY CAUTION FOR CONTIN...

Page 29: ...odel 7038 WARNING NO INTERNAL OPERATOR SERVICABLE PARTS SERVICE BY QUALIFIED PERSONNEL ONLY WARNING NO INTERNAL OPERATOR SERVICABLE PARTS SERVICE BY QUALIFIED PERSONNEL ONLY CAUTION FOR CONTINUED PROT...

Page 30: ...GAINST FIRE HAZARD REPLACE FUSE WITH SAME TYPE AND RATING CAUTION FOR CONTINUED PROTECTION AGAINST FIRE HAZARD REPLACE FUSE WITH SAME TYPE AND RATING INPUT 250V PEAK LINE RATING 50 60HZ 50VA MAX AC ON...

Page 31: ...o the bench or table Use only anti static type de soldering tools and grounded tip soldering irons 4 4 Principles of operation The following paragraphs discuss the basic Model 7038 op erating principl...

Page 32: ...105 perform this function The output enable OE lines of U101 U102 and U103 are controlled by the power up power down safeguard circuit located in the mainframe This circuit assures that no card relays...

Page 33: ...de the following information Card model number 7038 Card serial number Part description Circuit description if applicable Keithley part number 5 4 Factory service If the card is to be returned to Keit...

Page 34: ...108 RELAY HI FREQ RM2 5V RL 212 P2001 CONNECTOR 32 PIN 2 ROWS CS 775 1 Q100 Q102 Q104 TRANS NPN MMBT3904 TG 238 Q101 Q103 Q105 P CHANNEL ENHANCEMENT POWER FET TG 362 R100 R102 R103 R105 R106 R108 RES...

Page 35: ......

Page 36: ......

Page 37: ......

Page 38: ......

Page 39: ...2 Instruction manual 1 2 Input output connecting cables 2 3 Introduction 1 1 2 1 3 1 4 1 5 1 Isolation resistance tests 4 2 M Manual addenda 1 1 Matrix expansion 3 4 Multiplexer expansion 3 4 O Operat...

Page 40: ...nstable Will not read applied input Overload Calibration only Certi cate of calibration required Data required attach any additional sheets as necessary Show a block diagram of your measurement system...

Page 41: ...Keithley Instruments Inc 28775 Aurora Road Cleveland Ohio 44139 Printed in the U S A...