2.3.3 Input/output connecting cables
All connections to the multiplexer card input and output
jacks should be made using RG-179 75
Ω
coaxial cable
equipped with miniature SMB connectors.
WARNING
Make sure that power is off and external
energy sources are discharged before
connecting or disconnecting cables.
NOTE
75
Ω
cables must be used to assure good
high-frequency performance. See
Switching considerations
in
Section 3
for
more information.
2.3.4 Typical connecting scheme
Figure 2-3 shows a typical connecting scheme for the Model
7038. In this arrangement, sources are connected to the mul-
tiplexer A inputs while the measuring instrument is connect-
ed to the output.
CAUTION
Maximum voltage from any terminal
(center conductor or shield) to any other
terminal or chassis is 24V. Exceeding
this value may result in card damage.
IN1
IN2
IN3
IN4
OUTA
IN5
IN6
IN7
IN8
OUTB
IN9
IN10
IN11
IN12
OUTC
Note: All cables are 75
Ω
RG-179 recommended
Measuring Instrument
Figure 2-3
Typical connection scheme