7075 Eight 1
×
12 2-Pole Multiplexer Card
MULTIPLEX CONFIGURATION:
Eight 1
×
12 banks. Adjacent banks
can be connected together. Jumpers can be removed to isolate any
bank from the backplane.
CONTACT CONFIGURATION:
2-pole Form A (HI, GUARD).
CONNECTOR TYPE:
25-pin subminiature D connector, eight for
bank connection, one for row connection.
MAXIMUM SIGNAL LEVEL:
DC Signals:
110V DC pin to pin, 1A switched, 30VA (resistive
load).
AC Signals:
175V AC peak pin to pin, 1A switched, 60VA (resistive
load).
COMMON MODE VOLTAGE:
110V DC, 175V AC peak pin-to-pin or
pin-to-chassis.
CONTACT LIFE:
Cold Switching:
10
8
closures.
At Maximum Signal Level:
10
5
closures.
CHANNEL RESISTANCE (per conductor):
<0.50
Ω
initial, <1.5
Ω
at
end of contact life.
CONTACT POTENTIAL:
<5µV per contact pair (HI to GUARD).
OFFSET CURRENT:
<100pA.
CROSSTALK (1MHz, 50
Ω
load):
Bank: <
–60dB.
Channel:
<–60dB.
INSERTION LOSS (1MHz, 50
Ω
source, 50
Ω
load):
0.1dB typical.
ISOLATION:
Bank:
>10
10
Ω
, <3pF.
Channel
: >10
10
Ω
, <5pF.
Differential
:
Configured as 1
×
12: >10
9
Ω
, <100pF nominal.
Configured as 1
×
96: >10
8
Ω
, <600pF nominal.
Common Mode
:
Configured as 1
×
12: >10
9
Ω
, <165pF nominal.
Configured as 1
×
96: >10
8
Ω
, <700pF nominal.
3dB BANDWIDTH (50
Ω
load):
Configured as 1
×
12: 30MHz typical.
Configured as 1
×
96: 2.5MHz typical.
RELAY DRIVE CURRENT (per relay):
28mA.
RELAY SETTLING TIME:
<3ms.
EMC:
Conforms to European Union Directive 89/336/EEC.
SAFETY:
Conforms to European Union Directive 73/23/EEC (meets
EN61010-1/IEC 1010).
ENVIRONMENT:
Operating:
0° to 50°C, up to 35°C at 70% R.H.
Storage:
–25° to 65°C.
ACCESSORY SUPPLIED:
Jumpers for multiplexer expansion.
ACCESSORIES AVAILABLE:
7076-RMTC High Isolation Row Cable Assembly,
3m (10 ft)
7076-CMTC High Isolation Column/Bank Cable
Assembly, 3m (10 ft)
7075-MTC*
Row/Column/Bank Standard Cable
Assembly, 3m (10 ft)
* Standard cable can be used for either row or column connection.
A
B
C
D
E
F
G
H
1
2
3
4
5
6
7
8
9
10
11
12
ROWS
(To rear
panel
connectors)
BANKS—Eight 1
×
12
(To rear panel connectors)
Backplane
Jumpers
(factory installed)
Bank-to-
Bank
Jumpers
(user
installable)
Summary of Contents for 7075
Page 6: ......
Page 7: ......
Page 11: ......
Page 15: ......
Page 16: ......
Page 17: ......
Page 18: ......
Page 19: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ......
Page 24: ......
Page 25: ......
Page 26: ......
Page 27: ......
Page 28: ......
Page 29: ......
Page 30: ......
Page 31: ......
Page 32: ......
Page 33: ......
Page 34: ......
Page 35: ......
Page 36: ......
Page 37: ......
Page 38: ......
Page 39: ......
Page 40: ......
Page 41: ......
Page 42: ......
Page 43: ......
Page 44: ......
Page 45: ......
Page 46: ......
Page 47: ......
Page 48: ......
Page 49: ......
Page 51: ......
Page 52: ......
Page 53: ......
Page 54: ......
Page 55: ......
Page 56: ......
Page 57: ......
Page 58: ......
Page 59: ......
Page 60: ......
Page 61: ......
Page 62: ......
Page 63: ......
Page 64: ......
Page 65: ......
Page 66: ......
Page 67: ......
Page 68: ......
Page 69: ......
Page 70: ......
Page 71: ......
Page 72: ......
Page 73: ......
Page 74: ......
Page 75: ......
Page 76: ......
Page 77: ......
Page 78: ......
Page 79: ......
Page 80: ......
Page 81: ......
Page 82: ......
Page 83: ......
Page 84: ......
Page 85: ......
Page 86: ......
Page 87: ......
Page 88: ......
Page 89: ......
Page 90: ......
Page 91: ......
Page 92: ......
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ......
Page 97: ......
Page 98: ......
Page 99: ......
Page 100: ......
Page 101: ......
Page 102: ......
Page 103: ......
Page 104: ......
Page 105: ......
Page 106: ......
Page 107: ......