Operation
4-3
To connect and power up a master/slave configuration for the
first time, follow these steps:
1. Connect up to five mainframes in a daisy-chain (MAS-
TER/SLAVE OUT of one unit to MASTER/SLAVE IN
of next unit). Refer to paragraph 3.5.4 for more informa-
tion.
2. Power down each unit.
3. From the rear panel of the desired master unit, set the
IEEE-488 address to 30 plus the present value. This ac-
tion identifies the master unit while keeping the address
at the previous number, which is the master’s IEEE-488
setting minus 30.
4. Power up all the Model 708A Switching Systems.
During this initial and all subsequent power-ups of master/
slave configurations, connected Model 708A Switching Ma-
trices wait until all units have power before initializing. It is
not necessary to turn on the master unit last.
CAUTION
When it is necessary to cycle power on a
slave unit, turn off all units in the mas-
ter/slave configuration. This procedure
prevents the open communication and
control loop from putting the slave unit
in an undesirable state.
4.3.4 Power-up configuration
After the power-up tests and display messages are complet-
ed, the Model 708A assumes the following specific operating
states:
• All relays open.
• Crosspoint display shows present relay setup.
• Relay Step to 000 (a pseudo setup memory that is
cleared at powerup and sent to the relays).
• Memory Step to 001.
• Triggers are disabled.
Table 4-2 summarizes the power-up configuration for the
unit. The entire power-up process takes approximately five
seconds to complete.
4.3.5 Master/slave power-up
The power-up sequence for Model 708A mainframes can be
summarized as follows:
• Units previously programmed as stand-alones or slaves
power up as stand-alones.
• A unit previously programmed as a master powers up as
a master and tries to initiate a loop connection. If it is
successful, other units in the loop become slaves. If it is
not successful, the error LED (ERR) illuminates and the
unit reverts to stand-alone operation.
Table 4-2
Power-up, reset, and factory defaults
Parameter
Power-up/reset default
Factory default
Relays
Stored Setups
Relay Step
Memory Step
Digital Output
Digital Power
External Trigger
Matrix Ready
Master/Slave
IEEE-488 Address
Programmed Settling Time
Make/Break Rows
Break/Make Rows
Trigger Enable
Trigger Source
All opened
Unchanged
000
001
000
Unchanged (user selectable— Jumper
W101 selects internal/external power)
Falling edge
Active low
Unchanged (if successful)
Unchanged
0ms
Unchanged
Unchanged
Disabled
External
All opened
All cleared
000
001
000
Internal
Falling edge
Active low
Unchanged
18
0ms
None selected
None selected
Disabled
External
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for 708A
Page 135: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 136: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 137: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 138: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 139: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 145: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 146: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 147: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 148: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 149: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 150: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...