Operation
4-4
4.4
Display
4.4.1 Error LED
Errors are signaled when the error LED (ERR) illuminates.
Use the U1 — error status command (paragraph 5.9.20) to
retrieve the cause of the error. Table 4-3 lists Model 708A er-
ror conditions. Where applicable, the necessary corrective
action is also given in the table.
4.4.2 IEEE-488 status indicators
The TALK, LSTN, and REM LEDs shown in Figure 4-2 in-
dicate modes when the Model 708A is being programmed
over the IEEE-488 bus. The TALK and LSTN indicators
show when the unit has been addressed to talk or listen.
These talk and listen commands are derived from the unit’s
primary address. REM turns on to show when the unit is
placed in remote by addressing it to listen. All front panel
controls except LOCAL and POWER are inoperative when
REM is on. Local operation is restored by pressing LOCAL
unless the IEEE-488 LLO (local lockout) command is in ef-
fect. See Section 5 for details about IEEE-488 bus operation.
Table 4-3
Error conditions
Error
Description
Corrective action
Card ID Error*
IDDC
IDDCO
Invalid Input
M/S Error*
M/S Loop Down
Not in Remote
RAM Fail*
ROM Fail*
Setup Error*
Trig Overrun
Checksum test failed on a card.
Invalid device-dependent command.
Invalid device-dependent command option.
Invalid crosspoint address, setup location,
make/break or break/make row, or parame-
ter out of range.
Error in master/slave communication loop
(overrun, parity, framing, count imbalance,
or time-out).
One or more units connected in master/slave
loop are not powered up.
“X” character received over IEEE-488 bus,
but Model 708A is not in remote.
Self-test detected error in RAM.
Self-test detected checksum error in ROM.
Self-test detected checksum error in stored
setup. Battery may be low.
An additional trigger was received before
the Model 708A asserts the READY signal.
Remove card identified by all crosspoint LEDs lit.
Send only valid commands (see Section 5).
Send only valid command options (see Section 5).
Enter valid data.
Check for a closed loop of MASTER/SLAVE OUT
to MASTER/SLAVE IN.
Turn on all units or reconfigure master/slave loop.
Put Model 708A in remote.
See troubleshooting in Section 7.
See troubleshooting in Section 7.
Affected setup is cleared, and then Model 708A
proceeds normally.
Check the READY bit in the serial poll byte.
* Error LED (ERR) remains lit until next operation.
SYSTEM
COPY
12
TALK
LSTN
REM
ERR
COPY
PO
Figure 4-2
IEEE-488 status indicators
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Summary of Contents for 708A
Page 135: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 136: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 137: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 138: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 139: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 145: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 146: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 147: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 148: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 149: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...
Page 150: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...