Model 7701
21
Test procedure:
NOTE
The following test procedure assumes a Model 7701 switching module installed in slot 1 of the
mainframe.
Multiple channel operation from the front panel is not available for early versions of the
Model 2700.
1.
Open all channels.
Front panel operation: Press OPEN > Display ALL > Press OPEN
Remote programming:
ROUT:OPEN:ALL
2.
Close channel 33 to isolate measure channels (1 through 16) from source channels (17 through
32).
Front panel operation: Press CLOSE > Select MULTI > Key in 133 >
Press ENTER
Remote programming:
ROUT:MULT:CLOS (@133)
3.
Close channel 35 to connect measure channels (1 through 16) to DMM Input.
Front panel operation: Press CLOSE > Select MULTI > Key in 135 >
Press ENTER
Remote programming:
ROUT:MULT:CLOS (@135)
4.
Close channel 1 to connect DUT 1 to the DMM.
Front panel operation: Press CLOSE > Select MULTI > Key in 101 >
Press ENTER
Remote programming:
ROUT:MULT:CLOS (@101)
5.
Close channel 17 to connect DUT 1 to the external source.
Front panel operation: Press CLOSE > Select MULTI > Key in 117 >
Press ENTER
Remote programming:
ROUT:MULT:CLOS (@117)
6.
Measure DUT 1.
Front panel operation: Take reading from display
Remote programming:
DATA? (for continuous triggering mode)
READ? (for one-shot triggering mode)
7.
Open channels 1 and 17 to disconnect the DMM and external source from DUT 1.
Front panel operation: Press OPEN > Select MULTI > Key in 101 >
Press ENTER
Press OPEN > Select MULTI > Key in 117 >
Press ENTER
Remote programming:
ROUT:MULT:OPEN (@101,117)
8.
In general, repeat steps 4 through 7 to test DUT 2 through DUT 16. That is, close the channels to
connect the DUT to the DMM and external source, take a measurement, and then open the chan-
nels to disconnect the DUT from the DMM and external source.
9.
After all DUT is tested, repeat step 1 to open all channels.