Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-19
:ACAL:NEXTrun:TIME?
This command returns the date and time when the next auto calibration is scheduled to be run.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Nonvolatile memory
Not applicable
Usage
:ACAL:NEXTrun:TIME?
Details
The date and time is returned in the format:
MM/DD/YYYY HH:MM:SS.NNNNNNNNN
Where:
•
MM/DD/YYYY
is the month, date, and year
•
HH:MM:SS.NNNNNNNNN
is the hour, minute, second, and fractional second
Example
ACAL:NEXT:TIME?
Returns date and time when the next auto
calibration is scheduled to be run.
Example output:
05/29/2014 17:11:17.000000000
Also see
(on page 6-20)
:ACAL:REVert
This command returns auto calibration constants to the previous constants.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Nonvolatile memory
Not applicable
Usage
:ACAL:REVert
Details
This command reverts the present set of auto calibration constants to the previous set of auto
calibration constants.
The last run time and internal temperature are reverted to the previous values. The auto calibration
count is not changed.
Example
ACAL:REV
Auto calibration values are reverted to the
previous set of auto calibration constants.
Also see
(on page 6-20)