Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-69
Example 1
CURR:AVER:COUNT 10
CURR:AVER:TCON MOV
CURR:AVER ON
Set the averaging filter type to moving
average, with a filter count of 10.
Enable the averaging filter.
Example 2
RES:AVER:COUNT 10
RES:AVER:TCON MOV
RES:AVER ON
Set the averaging filter type to moving
average, with a filter count of 10.
Enable the averaging filter.
Example 3
VOLT:AVER:COUNT 10
VOLT:AVER:TCON MOV
VOLT:AVER ON
Set the averaging filter type to moving
average, with a filter count of 10.
Enable the averaging filter.
Also see
(on page 3-11)
[:SENSe[1]]:<function>:AVERage:COUNt
(on page 6-67)
[:SENSe[1]]:<function>:AVERage:TCONtrol
[:SENSe[1]]:<function>:AVERage:WINDow
(on page 6-71)
[:SENSe[1]]:<function>:AVERage:TCONtrol
This command sets the type of averaging filter that is used for the selected measure function when the
measurement filter is enabled.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
REP
Usage
[:SENSe[1]]:<function>:AVERage:TCONtrol <type>
[:SENSe[1]]:<function>:AVERage:TCONtrol?
<function>
The function to which the setting applies; see
Functions
<type>
The filter type to use when filtering is enabled; set to one of the following values:
•
Repeating filter:
REPeat
•
Moving filter:
MOVing
Functions
VOLTage[:DC]
RESistance
TEMPerature
VOLTage[:DC]:RATio
VOLTage:AC
FRESistance
CONTinuity
DIGitize:VOLTage
CURRent[:DC]
DIODe
FREQuency[:VOLTage]
DIGitize:CURRent
CURRent:AC
CAPacitance
PERiod[:VOLTage]