Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 6: SCPI command reference
DMM7510-901-01 Rev. B / May 2015
6-51
ROUTe subsystem
The ROUTe subsystem selects which set of input and output terminals to enable (front panel or rear
panel).
:ROUTe:TERMinals
This command describes which set of input and output terminals the instrument is using.
Type
Affected by
Where saved
Default value
Query only
Not applicable
Not applicable
Not applicable
Usage
:ROUTe:TERMinals?
Details
You must use the front-panel TERMINALS button to change which set of terminals the instrument
reads.
This query returns the set of input and output terminals that the instrument is using. If the instrument
is using the front-panel terminals, the return is:
FRON
If the instrument is using the rear-panel terminals, the return is:
REAR
Example
:ROUT:TERM?
Query to verify which terminals are used.
Output if the rear terminals are used:
REAR
Also see
None