Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 8: TSP command reference
DMM7510-901-01 Rev. B / May 2015
8-187
dmm.measure.rtdzero
This attribute contains the zero value of a user-defined RTD.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
100
Usage
value
= dmm.measure.rtdzero
dmm.measure.rtdzero =
value
value
The zero value of the RTD:
0
to
10000
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_RESISTANCE
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DC_CURRENT
dmm.FUNC_DIODE
dmm.FUNC_DCV_RATIO
dmm.FUNC_AC_CURRENT
dmm.FUNC_CAPACITANCE
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_TEMPERATURE
dmm.FUNC_CONTINUITY
dmm.FUNC_DIGITIZE_VOLTAGE
Details
This attribute is only valid when:
•
The function is set to temperature.
•
The transducer type is set to 3 or 4-wire RTD.
•
The RTD type is set to user-defined.
Example
dmm.measure.func = dmm.FUNC_TEMPERATURE
dmm.measure.transducer = dmm.TRANS_THREERTD
dmm.measure.threertd = dmm.RTD_USER
dmm.measure.rtdalpha = 0.00385
dmm.measure.rtdzero = 120
Set the measure function to temperature.
Set the transducer type to 3-wire RTD.
Set the RTD type to User.
Set the alpha RTD value to 0.00385.
Set the zero RTD value to 120.
Also see