Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 8: TSP command reference
DMM7510-901-01 Rev. B / May 2015
8-193
dmm.measure.threertd
This attribute defines the type of three-wire RTD that is being used.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
PT100
Usage
value
= dmm.measure.threertd
dmm.measure.threertd =
value
value
The type for 3-wire RTD:
•
PT100:
dmm.RTD_PT100
•
PT385:
dmm.RTD_PT385
•
PT3916:
dmm.RTD_PT3916
•
D100:
dmm.RTD_D100
•
F100:
dmm.RTD_F100
•
User-specified type:
dmm.RTD_USER
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_RESISTANCE
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DC_CURRENT
dmm.FUNC_DIODE
dmm.FUNC_DCV_RATIO
dmm.FUNC_AC_CURRENT
dmm.FUNC_CAPACITANCE
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_TEMPERATURE
dmm.FUNC_CONTINUITY
dmm.FUNC_DIGITIZE_VOLTAGE
Details
The transducer type must be set to temperature and the transducer must be set to 3-wire RTD before
you can set the RTD type.
Example
dmm.measure.func = dmm.FUNC_TEMPERATURE
dmm.measure.transducer = dmm.TRANS_THREERTD
dmm.measure.threertd = dmm.RTD_D100
Set the measure function to temperature.
Set the transducer type to 3-wire RTD.
Set the RTD type to D100.
Also see
(on page 8-184)
(on page 8-185)
(on page 8-186)
(on page 8-187)
(on page 2-120)