18 Integra Series
Quick Start Guide
TRIG:COUN 1
‘ Sets trigger count to one.
SENS:FUNC ‘TEMP’
‘ Select Frequency function.
SENS:TEMP:TRAN THER
‘ Select thermistor transducer.
SENS:TEMP:THER 5000
‘ Select 5000 ohm thermistor.
READ?
‘ Triggers measurement and requests one reading.
NOTE
When finished with remote programming, press the LOCAL key to
restore front panel control.
Measure frequency and period
The Model 27xx can measure frequency or period (Period = 1 / Frequency).
1.
Connections
– Connect the AC signal as shown in
A (b).
2.
Function
– Referring to
, select the Frequency or Period function:
•
Frequency – Press the FREQ key.
•
Period – Press the SHIFT key, then press FREQ (PERIOD).
3.
Select a measure range that will accommodate the expected AC input level.
4.
Reading
– Take a reading from the display.
SCPI programming commands (frequency and period)
Frequency
SENS:FUNC ‘FREQ’
‘ Select Frequency function.
SENS:FREQ:THR <NRf>
‘ Specify expected input level (in volts): <NRf> = 0 to 1010.
Period
SENS:FUNC ‘PER’
‘ Select Period function.
SENS:PER:THR <NRf>
‘ Specify expected input level (in volts): <NRf> = 0 to 1010.
Remote programming example (frequency)
INIT:CONT OFF
‘ Disables continuous initiation.
TRIG:COUN 1
‘ Sets trigger count to one.
SENS:FUNC ‘FREQ’
‘ Selects Frequency function.
SENS:FREQ:THR 10
‘ Selects 10V threshold range for expected input.
READ?
‘ Triggers measurement and requests one reading.
NOTE
When finished with remote programming, press the LOCAL key to
restore front panel control.