Ground, Power, and Isolation Connections
63
A
A
A
A
A
A
A
A
A
Keep the following in mind:
• Earth ground on the KUSB-3116 module is not connected to
DGND or AGND.
• Earth ground is connected to the aluminum case of the BNC
connection box.
• You should connect earth ground to the power supply earth.
• You should isolate the +5V/DGND input. Note that the power
supply (shipped with the KUSB-3116 module) has no connection
b5V/DGND and earth ground.
• The USB connector case is connected to earth ground.
• The USB data lines and USB GND are not connected to earth
ground.
• The USB DGND is connected to the USB GND of the PC USB
port.