S530 Parametric Test System Administrative Guide
Section 4: Maintenance
S530-924-01 Rev. D / September 2017
4-11
5.
At the prompt, enter
1
and press
Enter
to select the DMH (datahub) option.
6.
Follow the remaining prompts, and press
Enter
after each response you type.
7.
Once the installation has completed, type
$KIBIN/shutdown_S530.pl
and press
Enter
to
power down the S530 system.
8.
Power up the entire system.
9.
Log on as
kthmgr
.
To upgrade systems with KTE version 5.5.0 and later:
1.
Open a new Linux
®
terminal session.
2.
Type
licCheck
and press
Enter
to check if the datahub license is installed correctly.
Figure 45: Check datahub license installation
3.
If it is not installed correctly, you will need to edit the file at
opt/kiS530/.ki_setup
.
4.
Go to line 33 in the file and change it to
setenv COMPUTERNAME q1234
(where
1234
is your
QMO number).
5.
Go to line 61 in the file and change it to
setenv KI_QMO 1234
(where
1234
is your QMO
number).
6.
Reboot the computer to reinitialize the datahub.
LO patch panel
The S530 must have a single, direct connection between instrument lows and protective earth (safety
ground). Keithley recommends that a high-quality, low-impedance connection between low and
protective earth (safety ground) be made at the prober used with the S530. Because this is not
always possible, the LO patch panel provides alternative connections. The LO patch panel provides:
A common reference point for all the instrument low-side connections
In 4-wire systems, a common connection for the sense low terminals of the source-measure units
(SMUs)
Connections from low to sense low using a 100 kΩ resistor to enable autosensing
Different ways to tie low to protective earth (safety ground)
The LO patch panel is configured by the Keithley Field Service Engineer (FSE).