S530 Parametric Test System Administrative Guide
Section 3: Equipment startup
S530-924-01 Rev. D / September 2017
3-5
Shut down using KTE
To shut down using the Keithley Test Environment (KTE) software:
1.
Close all KTE programs.
2.
In the LINUX
®
terminal, type the following command:
$KIBIN/shutdown_S530.pl
3.
Press
Enter
on the keyboard.
4.
On the next command line, type the following:
sudo $KIBIN/shutdown_S530.pl
5.
Enter the password for the root account.
6.
Wait for the system to stop.
The IC process stops, the 4200-SCS shuts down, and then the workstation will shut down. The
console screen will be blank when it has stopped.
7.
Turn off power to the cabinet.
Emergency OFF (EMO) button
An EMERGENCY OFF (EMO) button is on the system cabinet door (see following figure). If you push
the EMERGENCY OFF button, it removes power to all of the system instruments. However, it does
not remove power to the host computer.
Figure 39: EMERGENCY OFF button
The EMO TRIPPED indicator light (on the cabinet door) turns on when the system has undergone an
emergency shut down.
Emergency shut down procedure
Press the red
EMERGENCY OFF
button on the front of the system cabinet. The instruments will
power down and the red EMO TRIPPED indicator will illuminate.