Performance Veri
fi
cation
8.
Change the AC Power Source output to the maximum voltage speci
fi
ed in the
following table.
DUT voltage selector
switch
AC Power Source voltage
Japan con
fi
gured units
110
132 V
110 V
220
264 V
220 V
9.
Enter the voltmeter reading into table for checking DC Voltage Line
Regulation under the Max column for the CUT. (See Table 18 on page 19.)
10.
Change the AC Power Source output to match the Voltage Selector switch
setting of the DUT.
DUT voltage selector
switch
AC Power Source voltage
Japan con
fi
gured units
110
115 V
100 V
220
230 V
200 V
11.
Enter the voltmeter reading into table for checking DC Voltage Line
Regulation under the Nom column for the CUT. (See Table 18 on page 19.)
12.
Calculate the two values: Nom – Min and Max – Nom. Enter the values into
the appropriate columns. (See Table 18 on page 19.)
13.
Enter the largest of the two values calculated in step 12 into the Largest
column. (See Table 18 on page 19.)
This completes the check for one channel. If needed, return to step 1 to run
through the check for the next channel.
36
Series 2200 Multichannel Programmable DC Power Supplies Technical Reference