2
STEP 4 Turning on the Output
16.
Press Output On/Off to turn on the output.
Remote Operation
The following SCPI command example enables channel 2, sets the voltage output to 2.22V, the current
limit to 1.5A and turns the output on.
INST:SEL CH2
SOURCE:OUTP:ENAB ON
SOURCE:VOLT 2.22V
SOURCE:CURR 1.5A
SOURCE:OUTP ON
Combining Channels 1 and 2 in Series to Output More than 30V
The series 2200 multi-‐channel power supplies have all channels isolated in their default state. However,
channel 1 and channel 2 can be combined in series to output a higher total voltage. The following
example demonstrates the steps to enable and output a combined voltage from channel 1 and 2.
Front Panel Operation
STEP 1 Configure the Instrument to Series Combination
1.
Press Menu.
2.
Use the arrow keys or the navigation wheel to select
Combine CH1+CH2
.
3.
Press Enter to enter the submenu.
4.
Use the arrow keys or the navigation wheel to select
V1+V2 Series
.
5.
Press Enter to set to series combination. See an example of a series combination front panel
display below.
STEP 2 Setting the Voltage Output and the Current Limit
6.
Press CH1.
7.
Press V-‐Set.
8.
Use the arrow keys, numeric keypad, or the navigation wheel to enter the voltage output.
9.
Press Enter to set the voltage of the combined channels.
10.
Press I-‐Set.
11.
Use the arrow keys, numeric keypad, or the navigation wheel to enter the current limit.
12.
Press Enter to set the current of the combined channel.
STEP 3 Turning on the Output
13.
Press Output On/Off to turn on the output.