background image

   

2   

STEP   4   Turning   on   the   Output   

16.

 

Press   Output   On/Off   to   turn   on   the   output.   

   

Remote   Operation   

   
The   following   SCPI   command   example   enables   channel   2,   sets   the   voltage   output   to   2.22V,   the   current   
limit   to   1.5A   and   turns   the   output   on.   
   

INST:SEL   CH2   
SOURCE:OUTP:ENAB   ON   
SOURCE:VOLT   2.22V   
SOURCE:CURR   1.5A   
SOURCE:OUTP   ON   

   

Combining   Channels   1   and   2   in   Series   to   Output   More   than   30V   

   
The   series   2200   multi-­‐channel   power   supplies   have   all   channels   isolated   in   their   default   state.      However,   
channel   1   and   channel   2   can   be   combined   in   series   to   output   a   higher   total   voltage.      The   following   
example   demonstrates   the   steps   to   enable   and   output   a   combined   voltage   from   channel   1   and   2.   
   

Front   Panel   Operation   

STEP   1   Configure   the   Instrument   to   Series   Combination   

1.

 

Press   Menu.   

2.

 

Use   the   arrow   keys   or   the   navigation   wheel   to   select   

Combine   CH1+CH2

.   

3.

 

Press   Enter   to   enter   the   submenu.   

4.

 

Use   the   arrow   keys   or   the   navigation   wheel   to   select   

V1+V2   Series

.   

5.

 

Press   Enter   to   set   to   series   combination.         See   an   example   of   a   series   combination   front   panel   
display   below.   
   

   

STEP   2   Setting   the   Voltage   Output   and   the   Current   Limit   

6.

 

Press   CH1.      

7.

 

Press   V-­‐Set.   

8.

 

Use   the   arrow   keys,   numeric   keypad,   or   the   navigation   wheel   to   enter   the   voltage   output.   

9.

 

Press   Enter   to   set   the   voltage   of   the   combined   channels.   

10.

 

Press   I-­‐Set.   

11.

 

Use   the   arrow   keys,   numeric   keypad,   or   the   navigation   wheel   to   enter   the   current   limit.   

12.

 

Press   Enter   to   set   the   current   of   the   combined   channel.   

STEP   3   Turning   on   the   Output   

13.

 

Press   Output   On/Off   to   turn   on   the   output.   

Summary of Contents for Series 2200

Page 1: ...o select Enable Channels 3 Press Enter to enter the submenu 4 Use the arrow keys or the navigation wheel to select Enable Disable CH2 5 Press Enter to enter the submenu 6 Use the arrow keys or the nav...

Page 2: ...channel 1 and 2 Front Panel Operation STEP 1 Configure the Instrument to Series Combination 1 Press Menu 2 Use the arrow keys or the navigation wheel to select Combine CH1 CH2 3 Press Enter to enter t...

Page 3: ...t S e n s e CHANNEL 2 O u t S e n e O u t S e n s e CHANNEL 3 O u t S e n s e REAR PANEL SOURCE HIGH SOURCE LOW SENSE HIGH SENSE LOW Figure 1 Connecting a Device in Series Combination on the Front Pa...

Page 4: ...limit 12 Press Enter to set the current of the combined channel STEP 3 Turning on the Output 13 Press Output On Off Remote Operation The following SCPI command example configures the power supply for...

Page 5: ...2 Press V Set 3 Use the arrow keys numeric keypad or the navigation wheel to enter a voltage output e g 10V 4 Press Enter to set the voltage of the combined channel 5 Press I Set 6 Use the arrow keys...

Page 6: ...will change the other channel by the desired ratio Remote Operation The following SCPI command example sets CH1 voltage output to 10V the current limit to 1 5A CH2 voltage output to track half of CH1...

Page 7: ...n the Front Panel and the Rear Panel Turning On Multiple Channels Sequentially The Series 2200 multi channel power supplies have all channels enabled in its default state The following example demonst...

Page 8: ...nable Disable CH1 26 Press Enter to enter the submenu 27 Use the arrow keys or the navigation wheel to select Enable CH1 default 28 Press Enter to enable channel 1 29 Press Esc a couple of times to re...

Page 9: ...e arrow keys or the navigation wheel to select Disable CH2 61 Press Enter to disable channel 2 and turn off channel 2 s output voltage 62 Press Esc a couple of times to return to the main display STEP...

Page 10: ...nnel 3 turn on using the appropriate command syntax of the programming environment INST SEL CH3 SOURCE OUTP ENAB ON perform tests on the load INST SEL CH3 SOURCE OUTP ENAB OFF insert desired amount of...