Getting Started
Menu descriptions (cont.)
Primary
Menu
Secondary
Menu
Description
>Output
Recall
Sets the power output On/Off state after power up. “On” restores the state to that
which was in use before the power was last turned off. “Off” disables this function and
con
fi
gures the output channel to power up in the off state.
NOTE.
Wait 3 seconds after changing the settings to allow the settings to be completely
stored before powering off the instrument
>Save Last
Recalls the operating parameters of the power supply after power up. It does this by
saving the last setting of operating parameters before you turn the instrument off and
then restoring the saved setting when you turn the instrument back on.
NOTE.
Wait 3 seconds after changing the settings to allow the settings to be completely
stored before powering off the instrument.
>Key Beep
Turns on or off a beeper sound when you push a button or press a key. “On” enables
the key sound. “Off” disables it.
>User
Settings
>Knob Lock
Locks the multipurpose knob.
>Address
Select a GPIB address for use with the Keithley GPIB to Series 2200 Multichannel Rear
Panel Mating Connector (CS-1655-15).
> Error Log
Lists all errors that have occurred.
Firmware-
/SW version
Shows the versions of
fi
rmware installed on the instrument.
>System Info
>Cal Date
Shows the date and time of the last calibration.
Channel status messages
Symbol
Description
OFF
The power supply output is off
CV
Constant voltage mode
CC
Constant current mode
Displayed error messages
Symbol
Description
EEPROM Test Fail
An EEPROM Error was detected during power-on self test. The instrument should be
returned to Keithley for service.
User Data Lost
A user data error was detected during power-on self test. Turn the instrument off and then
turn it back on to clear the error. If the error occurs repeatedly, the instrument should be
returned to Keithley for service.
Channel Initialization
Failed
An output channel failure was detected during power-on self test. The instrument should
be returned to Keithley for service.
Calibration Data Lost
A calibration data error was detected during power-on self test. Turn the instrument off and
then turn it back on to clear the error. If the error occurs repeatedly, the instrument should
be returned to Keithley for service.
Setup x Save Failure!
Saving a setup to memory x failed. The instrument should be returned to Keithley for service.
Series 2200 Multi-Channel Programmable DC Power Supplies User Manual