2400-902-01 (G - Feb 2006)
Model 2400 Service Manual
1-10
Performance Verification
Voltage measurement accuracy
Follow the steps below to verify that Model 2400 voltage measurement accuracy is within
specified limits. The test involves setting the source voltage to 95% of full-range values, as
measured by a precision digital multimeter, and then verifying that the Model 2400 voltage
readings are within required limits.
1.
With the power off, connect the digital multimeter to the Model 2400 INPUT/OUTPUT
jacks, as shown in Figure 1-1.
2.
Select the multimeter DC volts function.
NOTE
The default voltage source protection value is 40V. Before testing the 200V range, set
the voltage source protection value to >200V. To do so, press CONFIG then
SOURCE V to access the CONFIGURE V-SOURCE menu, then select PROTEC-
TION and set the limit value to >200V.
3.
Set the Model 2400 to both source and measure voltage by pressing the SOURCE V
and MEAS V keys, and make sure the source output is turned on.
4.
Verify output voltage accuracy for each of the voltages listed in Table 1-3. For each test
point:
•
Select the correct source range.
•
Set the Model 2400 output voltage to the indicated value as measured by the digital
multimeter.
•
Verify that the Model 2400 voltage reading is within the limits given in the table.
NOTE
It may not be possible to set the voltage source to the specified value. Use the closest
possible setting, and modify reading limits accordingly.
5.
Repeat the procedure for negative source voltages with the same magnitudes as those
listed in Table 1-3.
6.
Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-3
Voltage measurement accuracy limits
Model 2400 source and
measure range*
Source voltage**
Model 2400 voltage reading lim-
its (1 year, 18•C–28•C)
0
200mV
190.000mV
189.677 to 190.323mV
00
2V
1.90000V
1.89947 to 1.90053V
0
20V
19.0000V
18.9957to 19.0043V
200V
190.000V
189.962 to 190.038V
*
* Measure range coupled to source range when simultaneously sourcing and measuring voltage.
** As measured by precision digital multimeter. Use closest possible value, and modify reading limits
accordingly if necessary.
Summary of Contents for SourceMeter 2400
Page 10: ......
Page 13: ...1 PerformanceVerification...
Page 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Page 29: ...2 Calibration...
Page 51: ...3 RoutineMaintenance...
Page 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Page 55: ...4 Troubleshooting...
Page 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Page 69: ...5 Disassembly...
Page 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Page 77: ...6 ReplaceableParts...
Page 91: ...A Specifications...
Page 111: ...B CommandReference...
Page 123: ...C CalibrationPrograms...
Page 130: ...C 8 Calibration Programs...
Page 134: ......
Page 136: ......