Table 2-3
Front panel voltage calibration
Source range*
Source voltage
Multimeter voltage reading**
0.2V
2V
20V
200V
+200.00mV
+000.00mV
-200.00mV
-000.00mV
+2.0000V
+0.0000V
-2.0000V
-0.0000V
+20.000V
+00.000V
-20.000V
-00.000V
+200.00V
+000.00V
-200.00V
-000.00V
___________mV
___________mV
___________mV
___________mV
____________V
m
____________V
m
____________V
m
____________V
m
____________V
m
____________V
m
____________V
m
____________V
m
____________V
m
____________V
m
____________V
m
____________V
m
*
*Use EDIT and RANGE keys to select source range.
**Multimeter reading used in corresponding calibration step. See procedure.
2-11
Model 2400 Service Manual
Calibration
2400-902-01 (G - Feb 2006)
Summary of Contents for SourceMeter 2400
Page 10: ......
Page 13: ...1 PerformanceVerification...
Page 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Page 29: ...2 Calibration...
Page 51: ...3 RoutineMaintenance...
Page 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Page 55: ...4 Troubleshooting...
Page 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Page 69: ...5 Disassembly...
Page 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Page 77: ...6 ReplaceableParts...
Page 91: ...A Specifications...
Page 111: ...B CommandReference...
Page 123: ...C CalibrationPrograms...
Page 130: ...C 8 Calibration Programs...
Page 134: ......
Page 136: ......