2400-902-01 (G - Feb 2006)
Model 2400 Service Manual
3-2
Routine Maintenance
Introduction
The information in this section deals with routine type maintenance that can be performed by
the operator.
Line fuse replacement
WARNING
Disconnect the line cord at the rear panel, and remove all test leads con-
nected to the instrument (front and rear) before replacing the line fuse.
The power line fuse is accessible from the rear panel, just above the AC power receptacle (see
Figure 3-1).
Perform the following steps to replace the line fuse:
1.
Carefully grasp and squeeze together the locking tabs that secure the fuse carrier to the
fuse holder.
2.
Pull out the fuse carrier, and replace the fuse with the type specified in Table 3-1.
CAUTION
To prevent instrument damage, use only the fuse type specified in
Table 3-1.
3.
Reinstall the fuse carrier.
NOTE
If the power line fuse continues to blow, a circuit malfunction exists and must be
corrected. Refer to the troubleshooting section of this manual for additional
information.
Table 3-1
Power line fuse
Line voltage
Rating
Keithley part no.
100-240V
250V, 2.5A, slow blow 5
×
20mm
FU-106-2.5
Summary of Contents for SourceMeter 2400
Page 10: ......
Page 13: ...1 PerformanceVerification...
Page 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Page 29: ...2 Calibration...
Page 51: ...3 RoutineMaintenance...
Page 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Page 55: ...4 Troubleshooting...
Page 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Page 69: ...5 Disassembly...
Page 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Page 77: ...6 ReplaceableParts...
Page 91: ...A Specifications...
Page 111: ...B CommandReference...
Page 123: ...C CalibrationPrograms...
Page 130: ...C 8 Calibration Programs...
Page 134: ......
Page 136: ......