2400-902-01 (G - Feb 2006)
Model 2400 Service Manual
4-6
Troubleshooting
Power supply
Figure 4-2 shows a block diagram of the Model 2400 power delivery system.
The offline flyback switching power supply provides all power for the instrument while provid-
ing universal inputs for the 110/120V line. The digital board runs directly from the switcher,
including the +12VD supply. (See Digital circuitry.)
A constant-frequency switching supply runs off the +12VD supplies and generates all the float-
ing supply voltages for the analog board: +5VF, ±15VF, and ±30VF. An AC output (low volt-
age) supplies the analog board with the power it uses to derive the output stage supply voltages,
±36VO and ±220VO.
Analog Board
Output Stage
Constant Frequency
Low Noise Floating
Switching Supply
+12Vd
Line
Neutral
Switching Power
Supply
+12Vd
+5Vd
D
D
High Voltage/
Power
Digital Circuits
AC1
AC2
O
F
+30V
F
-30V
F
+15V
F
-15V
F
-220
+220
+36
+36
+5V
F
Figure 4-2
Power supply block diagram
Summary of Contents for SourceMeter 2400
Page 10: ......
Page 13: ...1 PerformanceVerification...
Page 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Page 29: ...2 Calibration...
Page 51: ...3 RoutineMaintenance...
Page 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Page 55: ...4 Troubleshooting...
Page 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Page 69: ...5 Disassembly...
Page 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Page 77: ...6 ReplaceableParts...
Page 91: ...A Specifications...
Page 111: ...B CommandReference...
Page 123: ...C CalibrationPrograms...
Page 130: ...C 8 Calibration Programs...
Page 134: ......
Page 136: ......