SPECIFICATIONS
SPEC-2400
A 18217
Released
SZ 1/11/96
B 18249
Revised
SZ 3/6/96
C 18608
Revised
SZ 4/22/96
D 18863
Revised
SZ 7/29/96
E 19739
Revised
SZ 10/4/98
F 27103
Revised
SZ 5/10/02
G 27953
Revised
SZ 3/04/03
H 29424
Revised
SZ 2/25/04
HW 2/25/04
Rev. H
Page 1 of 3
2400, 2400-LV and 2400-C SourceMeter
®
Specifications
SOURCE SPECIFICATIONS
1
VOLTAGE PROGRAMMING ACCURACY (LOCAL OR REMOTE SENSE)
MODEL RANGE
PROGRAMMING
RESOLUTION
ACCURACY (1 Year)
23°C
±
5°C
±
(% rdg. + volts)
NOISE
(peak-peak)
0.1Hz – 10Hz
200.00 mV
5 µV
0.02% + 600 µV
5 µV
2.00000 V
50 µV
0.02% + 600 µV
50 µV
20.0000 V
500 µV
0.02% + 2.4 mV
500 µV
2400-C and
2400 Only:
200.00 V
5 mV
0.02% + 24 mV
5 mV
TEMPERATURE COEFFICIENT (0°-18°C & 28°-50°C):
±
(0.15 × accuracy
specification)/
°C.
MAX. OUTPUT POWER:
22W, four quadrant source or sink operation.
SOURCE/SINK LIMITS:
MODEL 2400, 2400-C:
±
21V @
±
1.05A,
±
210V @
±
105 mA.
MODEL 2400-LV:
±
21V @
±
1.05A.
VOLTAGE REGULATION: Line:
0.01% of range.
Load:
0.01% of range + 100
µV
.
NOISE 10Hz – 1MHz (p-p):
10mV typical into a resistive load.
OVERVOLTAGE PROTECTION:
User selectable values, 5% tolerance. Factory
default = none.
CURRENT LIMIT:
Bipolar current limit (compliance) set with single value. Min.
0.1% of range.
OVERSHOOT:
<0.1% typical (full scale step, resistive load, 10mA range).
CURRENT PROGRAMMING ACCURACY (LOCAL OR REMOTE SENSE)
RANGE
PROGRAMMING
RESOLUTION
ACCURACY (1 Year)
3
23°C
±
5°C
±
(% rdg. + amps)
NOISE
(peak-peak)
0.1Hz – 10Hz
1.00000 µA
50 pA
0.035% + 600 pA
5 pA
10.0000 µA
500 pA
0.033% + 2 nA
50 pA
100.000 µA
5 nA
0.031% + 20 nA
500 pA
1.00000 mA
50 nA
0.034% + 200 nA
5 nA
10.0000 mA
500 nA
0.045% + 2 µA
50 nA
100.000 mA
5 µA
0.066% + 20 µA
500 nA
1.00000 A
2
50 µA
0.27% + 900 µA
5 µA
TEMPERATURE COEFFICIENT (0°-18°C & 28°-50°C):
±
(0.15 × accuracy
specification)/
°C.
MAX. OUTPUT POWER:
22W, four quadrant source or sink operation.
SOURCE/SINK LIMITS:
2400, 2400-C:
±
1.05A @
±
21V,
±
105 mA @
±
210V.
2400-LV:
±
1.05A @
±
21V.
CURRENT REGULATION: Line:
0.01% of range.
Load:
0.01% of range + 100
pA
.
VOLTAGE LIMIT:
Bipolar voltage limit (compliance) set with single value. Min.
0.1% of range.
OVERSHOOT:
<0.1% typical (1mA step, RL = 10k
Ω
, 20V range).
ADDITIONAL SOURCE SPECIFICATIONS
TRANSIENT RESPONSE TIME:
30
µs minimum for the output to
recover to its spec. following a step change in load
.
COMMAND PROCESSING TIME:
Maximum time required for the
output to begin to change following the receipt of
:SOURce:VOLTage|CURRent <nrf> command.
Autorange On:
10ms.
Autorange Off:
7ms.
OUTPUT SETTLING TIME:
Time required to reach 0.1% of final
value after command is processed. 100µs typical. Resistive load.
10µA to 100mA range.
OUTPUT SLEW RATE (±30%):
0.5V/µs, 200V range, 100mA
compliance. (2400 and 2400-C
only)
0.08V/µs, 20V range, 100mA
compliance.
DC FLOATING VOLTAGE:
Output can be floated up to ±250VDC
from chassis ground.
REMOTE SENSE:
Up to 1V drop per load lead.
COMPLIANCE ACCURACY:
Add 0.3% of range and ±0.02% of
reading to base specification.
OVER TEMPERATURE PROTECTION:
Internally sensed
temperature overload puts unit in standby mode.
RANGE CHANGE OVERSHOOT:
Overshoot into a fully resistive
100k
Ω
load, 10Hz to 1MHz BW, adjacent range changes between
200mV, 2V and 20V ranges, 100mV typical.
MINIMUM COMPLIANCE VALUE:
0.1% of range.
1. Specifications valid for continuous output currents below 105mA. For operation
above 105mA continuous for > 1 minute, derate accuracy 10%/35mA above
105mA.
2. Full operation (1A) regardless of load to 30°C. Above 30°C ambient, derate
35mA/°C and prorate 35mA/
Ω
load. 4-wire mode. For current sink operation on
1A range, maximum continuous power is limited to approximately 1/2 rated
power or less, depending on current, up to 30°C ambient. See power equations in
the User’s Manual to calculate allowable duty cycle for specific conditions.
3. For sink mode, 1µA to 100mA range, accuracy is:
±(0.15% + offset*4)
For 1A range, accuracy is:
±(1.5% + offset*8)
Summary of Contents for SourceMeter 2400
Page 10: ......
Page 13: ...1 PerformanceVerification...
Page 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Page 29: ...2 Calibration...
Page 51: ...3 RoutineMaintenance...
Page 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Page 55: ...4 Troubleshooting...
Page 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Page 69: ...5 Disassembly...
Page 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Page 77: ...6 ReplaceableParts...
Page 91: ...A Specifications...
Page 111: ...B CommandReference...
Page 123: ...C CalibrationPrograms...
Page 130: ...C 8 Calibration Programs...
Page 134: ......
Page 136: ......