Section 1: Introduction
Figure 1-7
Linear staircase sweep
Logarithmic staircase sweep
This sweep is similar to the linear staircase sweep. The steps, however, are done on a logarithmic
scale. The symmetrical log points for the steps are determined by the specified number of sweep
points. Figure 8 shows a 5-point log sweep from 1 to 10V.
With the desired source (V or I) selected, use the following menu sequence to set the start and
stop levels, and the number of sweep points (2 to 2500):
Press CONFIG > press SWEEP > select TYPE > select LOG
After selecting LOG, you will be prompted to enter the START and STOP levels, and the NO OF
POINTS.
Delay
Delay
Delay
Delay
Start
Bias
Stop
Step
Step
Step
Measure
Measure
Measure
Measure
X = Measurement point
X
X
X
X