Model 2460 Interactive SourceMeter® Instrument User's Manual
Section 3: Using a remote interface
2460-900-01 Rev. A / September 2014
3-3
To avoid possible mechanical damage, stack no more than three connectors on any one instrument.
To minimize interference caused by electromagnetic radiation, use only shielded GPIB cables.
Contact Keithley Instruments for shielded cables.
Figure 10: Model 2460 IEEE-488 connections
Set the GPIB address
The default GPIB address is 18. You can set the address to any address from 0 to 30 if it is unique in
the system. This address cannot conflict with an address that is assigned to another instrument or to
the GPIB controller.
GPIB controllers are usually set to 0 or 21. To be safe, do not configure any instrument to have an
address of 0 or 21. To change the controller address, see the documentation for the controller.
The instrument saves the address in nonvolatile memory. It does not change when you send a reset
command or when you turn the power off and then on again.