Model 2470 High Voltage SourceMeter Instrument User's Manual
Section 5: Leakage current and insulation resistance
2470-900-01 Rev. A /
May
2019
5-13
Set up the application using TSP commands
The following TSP code is designed to be run from Keithley Instruments Test Script Builder (TSB).
TSB is a software tool that is available from
. You can install and use TSB to write
code and develop scripts for TSP-enabled instruments. Information about how to use TSB is in the
online help for TSB and in the “Introduction to TSP operation” section of the
Model 2470
Reference Manual
.
By default, the 2470 uses the SCPI command set. You must select the TSP command set before
sending TSP commands to the instrument.
To enable TSP commands:
1.
Press the
MENU
key.
2.
Under System, select
Settings
.
3.
Set the Command Set to
TSP
.
4.
At the prompt to reboot, select
Yes
.
The following TSP commands make insulation resistance measurements by sourcing 700 V and
measuring the resistance. The Simple Loop trigger model template is used to make
10 measurements at 100 ms intervals. After the code is executed, the measurement results are
displayed in the Instrument Console of Test Script Builder.