status.operation.instrument.*
status.operation.instrument.smuX.condition
status.operation.instrument.smuX.enable
status.operation.instrument.smuX.event
status.operation.instrument.smuX.ntr
status.operation.instrument.smuX.ptr
* = condition, enable, event, ntr or ptr
smuX = smua or smub
Attribute
Operation SMU event register sets.
Usage
Reads condition, enable, event, NTR and PTR registers:
operreg = status.operation.instrument.smuX.condition
operreg = status.operation.instrument.smuX.enable
operreg = status.operation.instrument.smuX.event
operreg = status.operation.instrument.smuX.ntr
operreg = status.operation.instrument.smuX.ptr
Writes to enable, NTR and PTR registers:
status.operation.instrument.smuX.enable = operreg
status.operation.instrument.smuX.ntr = operreg
status.operation.instrument.smuX.ptr = operreg
Set
operreg
to one of the following values:
0
Clears all bits.
status.operation.CALIBRATING
Sets CAL bit (B0).
status.operation.CAL
Sets CAL bit (B0).
status.operation.MEASURING
Sets MEAS bit (B4).
status.operation.MEAS
Sets MEAS bit (B4).
status.operation.PROMPTS
Sets PRMPTS bit (B11).
status.operation.PRMPTS
Sets PRMPTS bit (B11).
status.operation.USER
Sets USER bit (B12).
status.operation.PROGRAM_RUNNING
Sets PROG bit (B14).
status.operation.PROG
Sets PROG bit (B14).
operreg
can also be set to the decimal weight of the bit to be set. Examples:
To set bit B0 (CAL), set
operreg
to 1 (2
0
).
To set bit B4 (MEAS), set
operreg
to 16 (2
4
).
To set bit B11 (PRMPTS), set
operreg
to 2048 (2
11
).
To set more than one bit of the register, set
operreg
to the sum of their decimal
weights. For example, to set bits B0 and B4, set
operreg
to 17 (1 + 16).