Function:
PulseIMeasureV(smu, bias, level, ton, toff, points)
Description: Performs a specified number of pulse I, measure V cycles:
• Sets the
smu
to output
bias
amps and dwell for
ton
seconds.
• Sets the
smu
to output
level
amps and dwell for
ton
seconds.
• Performs voltage measurement with source at
level
amps.
• Sets the
smu
to output
bias
amps for
toff
seconds.
• Repeats the above sequence for
points
pulse-measure cycles.
level
bias
ton
bias
ton
toff
Voltage measurement
Parameters:
smu, bias, level, ton, toff, points
smu:
SourceMeter Channel (A or B). Defaults to smua if all parameters
are omitted when function is called.
bias:
Bias level in amps.
level:
Pulse level in amps.
ton:
Pulse on-time in seconds.
toff:
Pulse off-time in seconds.
points:
Number of pulse-measure cycles.
Data: Pulsed voltage measurements, current levels and timestamps are stored in
smuX.nvbuffer1
.
Example:
PulseIMeasureV(smua, 0.001, 1.0, 20E-3, 40E-3, 10)
SMU A will output 1mA and dwell for 20ms, output 1A and dwell for 20ms,
and then perform a voltage measurement. After the measurement, the output
will return to 1mA and dwell for 40ms. This pulse-measure process will repeat
9 more times.