Series 2600 System SourceMeters Reference Manual
DUT Test Connections 3-7
Return to
2600S-901-01 Rev. A / May 2006
Remote programming sense selection
summarizes the commands to select the sense mode. See
for details on using these commands.
Table 3-2
Commands to select sense mode
Command
*
Description
smuX.source.output = smuX.OUTPUT_OFF Turns off the SMU output.
smuX.sense = smuX.SENSE_LOCAL
Selects local (2-wire) sense.
smuX.sense = smuX.SENSE_REMOTE
Selects remote (4-wire) sense.
* Model 2601/2611: smuX = smua.
Model 2602/2612: smuX = smua (Channel A) or smub (Channel B).
Contact check connections
The contact check function
1
prevents measurement errors due to excessive
resistance in the force or sense leads. Connections for contact check
measurements are shown in
for operation and
for details on contact check commands.
1. All Model 2611/2612s manufactured by Keithley Instruments support the contact check function. Only Model
2601/2602s with firmware revision 1.1.0 or later and SMU hardware revision E or later support the contact check
function. To determine the firmware and SMU hardware revisions, inspect the data returned by the
print(localnode.info()) command. The InstFwRev and SMUBrdRev keys contain the necessary information.
Summary of Contents for SourceMeter 2600 Series