Series 2600 System SourceMeters Reference Manual
Basic Operation 4-21
Return to
2600S-901-01 Rev. A / May 2006
Remote ohms programming
The following paragraphs summarize basic commands necessary for remote
ohms programming and also give a programming example for a typical ohms
measurement situation.
Remote ohms command
Use the following command to obtain a resistance reading:
reading = smuX.measure.r()
See
for more commands necessary to set up source and measure
functions, and also
Ohms programming example
The command sequence for a typical ohms measurement is shown below. These
commands set up the SourceMeter as follows:
• Source function: current, 10mA range, 10mA output
• Voltage measure range: auto
• Voltage compliance: 10V
• Sense mode: 4-wire
smua.reset()
--Restore Series 2600 defaults.
smua.source.func = smua.OUTPUT_DCAMPS
--Select current source function.
smua.source.rangei = 10e-3
--Set source range to 10mA.
smua.source.leveli = 10e-3
--Set current source to 10mA.
smua.source.limitv = 10
--Set voltage limit to 10V.
smua.sense = smua.SENSE_REMOTE
--Enable 4-wire ohms.
smua.measure.autorangev = smua.AUTORANGE_ON
--Set voltage range to auto.
smua.source.output = smua.OUTPUT_ON
--Turn on output.
print(smua.measure.r())
--Get resistance reading.
smua.source.output = smua.OUTPUT_OFF
--Turn off output.