Series 2600 System SourceMeters Reference Manual
Instrument Control Library 12-69
Return to
2600S-901-01 Rev. A / May 2006
smuX.measure.autorangeY
X = SMU channel (a or b)
Y = SMU measure function (v or i) where v = voltage, i = current
Attribute
Measure auto range setting.
Usage
autorange = smuX.measure.autorangeY
-- Reads measure auto range.
smuX.measure.autorangeY = autorange
-- Writes measure auto range.
Set
autorange
to one of the following values:
0
or
smuX.AUTORANGE_OFF
Disables measure auto range.
1
or
smuX.AUTORANGE_ON
Enables measure auto range.
Remarks
• This attribute indicates the measurement auto range state. Its value will be
smuX.AUTORANGE_OFF
when the SMU measure circuit is on a fixed range and
smuX.AUTORANGE_ON
when it is in auto range mode.
• Setting this attribute to
smuX.AUTORANGE_OFF
puts the SMU on a fixed range. The
fixed range used will be the range the SMU measure circuit was currently using.
• Setting this attribute to
smuX.AUTORANGE_ON
puts the SMU measure circuit into auto
range mode. It will remain on its present measure range until the next measurement
is requested.
Details
See “
Also see
Example
Enables voltage measure autoranging for SMU A:
smua.measure.autorangev = smua.AUTORANGE_ON
smuX.measure.autozero
X = SMU channel (a or b)
Attribute
Behavior of the SMU’s A/D internal reference measurements (autozero).
Usage
azmode = smuX.measure.autozero
-- Reads autozero.
smuX.measure.autozero = azmode
-- Writes autozero.
Set
azmode
to be one of the following values:
0
or
smuX.AUTOZERO_OFF
Autozero disabled.
1
or
smuX.AUTOZERO_ONCE
Performs autozero once, then disables autozero.
2
or
smuX.AUTOZERO_AUTO
Automatic checking of reference and zero
measurements. An autozero is performed when
needed.