status.system2.*
status.system2.condition
status.system2.enable
status.system2.event
* = condition, enable or event
Attribute
TSP-Link system2 data structure register set.
Usage
Reads condition, enable and event registers:
enablereg = status.system2.condition
enablereg = status.system2.enable
enablereg = status.system2.event
Writes to enable register:
status.system2.enable = enablereg
Set
enablereg
to one of the following values:
0
Clears all bits.
1
or
status.system2.EXTENSION_BIT
Sets EXT bit (B0).
1
or
status.system2.EXT
Sets EXT bit (B0).
status.system2.NODEn
Sets a node bit (Bn);
n
= 15 to 28.
Remarks
• In an expanded system (TSP-Link), this attribute is used to read or write to the
system2 node registers.
• Reading a system2 node register returns a numeric value whose binary equivalent
indicates which register bits are set. The bits of the system2 node register are
identified as follows:
Bit B0 – EXT bit
Bit B4 – Node 18
Bit B8 – Node 22
Bit B12 – Node 26
Bit B1 – Node 15
Bit B5 – Node 19
Bit B9 – Node 23
Bit B13 – Node 27
Bit B2 – Node 16
Bit B6 – Node 20
Bit B10 – Node 24
Bit B14 – Node 28
Bit B3 – Node 17
Bit B7 – Node 21
Bit B11 – Node 25
Bit B15 – Not used
For example, assume value 9 is returned for the enable register. The binary
equivalent is 0000000000001001. This value indicates that bit 0 (EXT) and bit 3
(Node 17) are set.
• Assigning a value to the
status.system2.enable
attribute sets the extension bit or
a node bit of the system2 node enable register.
Details
See “
System Summary Event Registers
Also see
Example
Sets the extension bit of the system2 enable register:
status.system2.enable = status.system2.EXT